As Needed Tasks
6.1 Program a New Test
4
Select
Menu List > Parameters > Specific Test Parameters
to
display the
General
tab.
•
Select
Edit (F1)
.
•
Select the test from the "Test Name" drop-down list.
•
Select
Serum
,
Urine
,
Other-1
or
Other-2
from the "Type" drop-
down list. Verify
Yes
is selected for the sample type for the new
test from the "Operation" drop-down list.
•
Enter the specific test parameters from the parameter sheet.
•
Verify the information and select
Confirm (F1)
.
TIP
Display the parameters for a maximum of 6 tests at a time for
verification. Select
List Display (F7)
. The first 6 tests display.
Use the arrows to page to the new test.