Daily Start Up
3.2 Perform Daily Maintenance
3.2 Perform Daily Maintenance
Inspect and perform Daily Maintenance to ensure system performance and
safety.
•
Inspect the syringes for leaks
•
Inspect, clean and prime the sample probes, reagent probes, and mix bars
•
Inspect the printer (optional) and paper
•
Replace the deionized (DI) water in the pre-dilution bottles
•
Prepare the probe wash solution
•
Inspect the stability of the upper cover
For details on performing daily maintenance inspections, refer to 8.3 Daily
Maintenance in the AU5800 User's Guide, chapter 8.
CAUTION
To prevent injury or infection be sure to wear protective gloves and
other protective equipment when performing maintenance.
For additional protection, observe the WARNING and CAUTION
statements in the
AU5800 User's Guide
.
3.3 Check Analyzer Status
Check
Analyzer Status
to display a color-coded overview of the system, and
the temperatures of the incubators and refrigerated R1/R2 compartments. The
rack set trays, rack collection trays, rack transport belts, wash solution tanks,
DI water tanks, waste tanks, printer, and LIS communication are monitored.
When the ISE is installed, a diagram of the ISE unit is also displayed.
Color
Status
Blue
No errors
Yellow or Orange
Non-fatal error, the unit can be started
Red
Fatal error, the unit cannot be started