Analyzer Modes
7.5 Perform an Emergency Stop
7
1
Press the
EM STOP
button. All power to the analyzer and ISE unit(s)
turns off immediately. The computer remains On. To turn off the
computer, press
(
Ctrl
)
+
(
ALT
)
+
(
Delete
)
. A Windows Security window
opens. Select the
Shutdown
button.
2
Remove all racks from the rack transport belts.
7.5.1 Return to
Standby
Mode
1
Press the
RESET
button.
2
Press the
ON
button. (The lamp(s) turn on and the software loads.)
3
In the New Index window, select
Current Index
to continue analysis
in the current index.
4
The system stays in Warm Up mode for 1.5 hours. After the required
20-minute lamp warm up time, select
Home > Analyzer
Maintenance > Stand By (F4)
to return to Standby.
5
Perform a W1.
For details on performing the W1, refer to 8.8.13 Perform a W1 Procedure in
the AU5800 User's Guide, chapter 8.