5-10
Maintenance
1. Close the Airfuge door and turn the instrument on its side.
2. Turn the time-delay knob counterclockwise to shorten the delay
time, or clockwise to lengthen the delay time.
3. Reconnect the power cord and the air supply and perform a trial
run to check the new setting.
DECONTAMINATION
If the instrument and/or accessories are contaminated with radioactive
or pathogenic solutions, follow appropriate decontamination
procedures as determined by your laboratory safety officer. Refer
to
Chemical Resistances
(publication IN-175) or contact Beckman
Coulter Field Service to ensure that the decontamination method
does not damage any part of the instrument (or accessories).
STERILIZATION AND DISINFECTION
Instrument surfaces are finished with vinyl paint, the window is made
of polycarbonate, and the front panel is acrylic. Ethanol (70%)* may
be used on these surfaces. See
Chemical Resistances
for more infor-
mation regarding chemical resistances of instrument and accessory
materials.
While Beckman Coulter has tested these methods and found that they
do not damage the instrument, no guarantee of sterility or disinfection
is expressed or implied. When sterilization or disinfection is a
concern, consult your laboratory safety officer regarding proper
methods to use.
* Flammability hazard. Do not use in or near operating ultracentrifuges.
Содержание Airfuge
Страница 1: ...AF IM 13 Instruction Manual Airfuge Air Driven Ultracentrifuge...
Страница 5: ...AF IM 13 December 2007 Instruction Manual Airfuge Air Driven Ultracentrifuge 2007 Beckman Coulter Inc...
Страница 6: ......
Страница 10: ......
Страница 30: ......
Страница 38: ......
Страница 50: ......
Страница 52: ......
Страница 53: ......
Страница 54: ......
Страница 55: ......