2-3
Preinstallation Requirements
FILTER INSTALLATION
➠
NOTE
If the air source is especially wet or oily, install a
second filter as a prefilter. A second filter will
cause an additional pressure drop of approxi-
mately 5 psig (35 kPa) in the air pressure that
reaches the instrument.
Make sure the required operating air pressure is obtainable at the
instrument PRESSURE gauge.
1. Pull back the spring-loaded sleeve of the quick-disconnect at the
filter outlet and push the fitting of the quick-disconnect hose into
the sleeve until they snap together (see Figure 2-1).
➠
NOTE
Do not install the filter upside down
. The filter
bottom has a valve that distinguishes it from the
top. Escaping air can be heard if the filter is
installed upside down.
Figure 2-1. Filter Installation
Filter Bottom
Remove barb fitting
before installing 1/4"
piping directly into filter
Free Standing Filter
Arrow indicates
direction of airflow
Quick-disconnect
hose from filter
to Airfuge
Ultracentrifuge
From Air
Source
Filter Mounted On The Wall
Quick-disconnect
hose from filter
to Airfuge
Ultracentrifuge
Arrow indicates
direction of airflow
Mounting
Screws (4)
Barb
Fitting
From Air Source
Содержание Airfuge
Страница 1: ...AF IM 13 Instruction Manual Airfuge Air Driven Ultracentrifuge...
Страница 5: ...AF IM 13 December 2007 Instruction Manual Airfuge Air Driven Ultracentrifuge 2007 Beckman Coulter Inc...
Страница 6: ......
Страница 10: ......
Страница 30: ......
Страница 38: ......
Страница 50: ......
Страница 52: ......
Страница 53: ......
Страница 54: ......
Страница 55: ......