1-6
Description
BRAKE PIN
Braking action is provided by a movable brake pin (see Figure 1-5)
set vertically on a spring in the shaft of the stator (see Figure 1-3).
An electromagnet holds the pin down during a run and releases it
2 minutes after the timer reaches zero. When released, the pin
engages the rotor bottom, stabilizing the rotor as it gradually
decelerates to a stop.
DOOR LATCH
The door-latch mechanism secures the door during centrifugation.
This prevents premature door opening, which would result in damage
to the rotor and stator pad.
FILTER
A filter is supplied with the ultracentrifuge for water and oil removal
from the air supply. The filter element turns red when it is saturated
and needs to be replaced.
NAME RATING PLATE
A name rating plate is affixed to the rear of the instrument. Always
mention the serial number and model number when contacting
Beckman Coulter regarding your Airfuge ultracentrifuge.