Parameterization and programming
ELX3312 and ELX3314
59
Version: 1.3.0
5.3.5
Limit 1 and Limit 2
Limit 1 and limit 2, index 0x80n0:13, index 0x80n0:14
A temperature range can be set that is limited by the values in the indices
. If the limit values are overshot, the bits in indices 0x60n0:03 and 0x60n0:05 are set.
The temperature value is entered with a resolution of 0.1°C.
Example
Limit 1= 30°C
Value index
= 300
5.3.6
Calibration
Vendor calibration, index 0x80n0:0B
The vendor calibration is enabled via index
. Parameterization takes place via the indices:
•
: Vendor offset calibration thermocouple
•
: Vendor gain calibration thermocouple
•
: Vendor CJ offset calibration
Vendor and user calibration
User calibration (index
) should only be performed instead of the vendor calibra-
tion (index
), but this is generally only necessary in exceptional cases.
User scaling, index 0x80n0:01
The user scaling is enabled via index
. Parameterization takes place via the indices
•
The offset describes a vertical shift of the characteristic curve by a linear amount.
At a resolution of 0.1°, 1 digit
(dec)
corresponds to an increase in measured value by 0.1°
At a resolution of 0.01°, 1 digit
(dec)
corresponds to an increase in measured value by 0.01
•
The default value of 65536
(dec)
corresponds to gain = 1.
The new gain value for 2-point user calibration after offset calibration is determined as follows:
Gain_new = reference temperature / measured value x 65536
(dec)
User calibration, index 0x80n0:0A
User calibration is enabled via index
. Parameterization takes place via the indices
•
: User offset calibration thermocouple
•
: User gain calibration thermocouple
Calculation of process data
Calibration
The concept
calibration
, which has historical roots at Beckhoff, is used here even if it has nothing to
do with the deviation statements of a calibration certificate. Actually, this is a description of the ven-
dor or customer calibration data/adjustment data used by the device during operation in order to
maintain the assured measuring accuracy.