4-13
MEGA3 NFPA Medical Gas Notification System
4107 9016 59.03
Figure 75: Gas Choices
Figure 76: Master Alarm User Interface
• Touch the Save button when you see a
check beside your selection.
• (If an alarm sounds, press the Mute but-
ton on the panel under the screen and
continue with setup.)
• Repeat to add additional gases.
• Touch the X button in the upper right to
close the GAS SENSORS screen and return
to the PRIMARY SETTINGS screen. The
alarm will retain your information.
• Touch the X button in the upper right to
close the PRIMARY SETTINGS screen. The
alarm will retain your information.
4.4 Setting Up a Master Alarm
The master alarm monitors whether source
equipment is working properly. Two indepen-
dent hard wired alarms are required per facil-
ity to meet the 24-hour monitoring require-
ment in the NFPA 99 standard.
The information for the signals connected in
the field needs to be entered into the alarm
using the touch screen interface.
To set up or change the configuration of an
alarm, follow the steps outlined in the initial
set-up for an unconfigured master alarm in
section 4.4.