MEGA3 NFPA Medical Gas Notification System
4107 9016 59.03
Figure 57: Gas Name Tab
Figure 58: Area Monitored Tab
Figure 59: Units of Measure Tab
Figure 60: Low Alarm Settings Tab
C = [Gas Name] Tab
Use the keyboard to change the gas name.
Touch the [<ÁÑ] to toggle to symbols and ex-
tended Latin Characters. Touch the [1AZ] key
to toggle back to letters and numbers.
D = Area Monitored Tab
E = [Units of Measure] Tab
To add locations to be monitored:
1. Touch an unused location tab.
2. Using the keyboard, enter the name of the
location that you would like to monitor
(e.g., ROOM 1). Enter multiple locations
as needed (e.g., Room 1, Room 2, Room
5).
3. A check appears when the location is se-
lected, the save button has to be pressed
if the selection is new.
To edit the location information, touch the
Pencil button.
To change the unit of measure:
1. Touch the tab for the appropriate unit of
measure.
2. Touch the Save button to close the screen.
F = Low Alarm Settings Tab
You can remove the low alarm set point or
customize the low alarm set point.
To remove the low alarm set point:
1. Touch the NONE tab.
2. Touch the check on the CONFIRM SET-
TINGS CHANGE screen.
To customize the low alarm set point:
1. Touch the ADJUST tab.
2. Enter the new set point.
3. Touch the Save button.