EMC Information-Guidance and Manufacture’s Declaration
CAUTION:
- The Infrared Thermometer JPD-FR412 needs special precautions regarding
EMC and needs to be installed and put into service according to the EMC
information provided for in the ACCOMPANYING DOCUMENTS.
- Portable and mobile RF communications equipment can affect Infrared
Thermometer JPD-FR412.
- The Infrared Thermometer JPD-FR412 should not be used adjacent to or
stacked with other equipment.
Guidance and manufacturer´s declaration – Electromagnetic emission –for all
equipment and systems
Guidance and manufacturer’s declaration – Electromagnetic
immunity –for all equipment and systems
Guidance and manufacturer´s declaration – Electromagnetic emission
The Infrared Thermometer JPD-FR412 is intended for use in the electro-
magnetic environment specified below. The customer or the user of the
Infrared Thermometer JPD-FR412 should assure that it is used in such an
environment.
Emissions
test
Compliance
Electromagnetic environment - guidance
RF
emissions
CISPR 11
Group 1
The Infrared Thermometer JPD-FR412
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
RF
emissions
CISPR 11
Class B
The Infrared Thermometer JPD-FR412 is
suitable for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Guidance and manufacturer´s declaration – Electromagnetic immunity
The Infrared Thermometer JPD-FR412 is intended for use in the electro-
magnetic environment specified below. The customer or the user of the
Infrared Thermometer JPD-FR412 should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment- guidance
Guidance and manufacturer´s declaration – Electromagnetic immunity
The Infrared Thermometer JPD-FR412 is intended for use in the electro-
magnetic environment specified below. The customer or the user of the
Infrared Thermometer JPD-FR412 should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment- guidance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
Portable and mobile RF
communications
equipment should
be used no closer
to any part of the
JPD-FR412,including
cables, than the recom-
mended separation dis-
tance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended
separation distance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6kV
contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
Power fre-
quency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Guidance and manufacturer´s declaration – Electromagnetic
immunity –for equipment and systems that are not life-supporting
d=
P 80 MHz to 800 MHz
3.5
[
[
E.
1
d=
P 800 MHz to 2.5 GHz
7
[
[
E.
1