BD Alaris™ neXus GP Volumetric Pump
Configuration
BDDF00558 Issue 5
53/72
Large Volume Pump Configuration Settings
AIL Limit
The single bubble Air-in-line alarm setting .
Bolus Mode
Controls whether or not the Pump allows bolus delivery method .
Bolus Rate Default
The default value for bolus rates .
Bolus Rate Max
The maximum permissible bolus rate .
Bolus Volume Max
The maximum permissible bolus volume .
Occlusion Alarm Pressure
The default occlusion limit .
Pressure Max
The maximum permitted occlusion limit .
Infusion Rate Max
The maximum permissible infusion rate .
KVO Rate
The maximum infusion rate when performing KVO delivery .
Near End of Infusion Point
The point defined in terms of remaining time at which the infusion is deemed to be
Near end
i .e . nearly complete .
Primary VTBI Max
The maximum VTBI for primary infusions .
Prime Rate
The rate at which priming is performed .
Prime Volume Max
Determines the maximum volume that can be purged as part of a given
prime operation .
Secondary Infusion
Allows the use of a secondary infusion (Piggyback) in the same channel .
Sec. Infusion Rate Max
The maximum permissible infusion rate for secondary infusions .
Sec. VTBI Max
The maximum permissible setting for the Volume To Be Infused for secondary
infusions .