BD Alaris™ neXus GP Volumetric Pump
Overview
BDDF00558 Issue 5
9/72
Main Display Features
Main Display – If VTBI has not been set (flow sensor must be used)
ON HOLD
RATE
VOLUME
VTBI
ml/h
VOLUME
ml
25.0
50
.0
WiFi
Infusion Status /
Drug Name /
Profile Name /
Primary or Secondary
(Only if secondary is
enabled in the data set)
Infusion Rate
Volume Infused
Softkeys
Softkey Identifiers
Clear Volume Infused Set VTBI Option
a)
ON HOLD
RATE
SET RATE WITH
ON HOLD
RATE
RATE TOO LOW
ON HOLD
RATE
RATE TOO HIGH
If the rate has not been set and is
showing at 0 .0ml/h, then message
a)
will be displayed .
b)
ON HOLD
RATE
SET RATE WITH
ON HOLD
RATE
RATE TOO LOW
ON HOLD
RATE
RATE TOO HIGH
If programmed rate is between
0 .0ml/h and 0 .1ml/h exclusive in
drug protocol, message
b
) will be
shown .
Main Display – If VTBI is set
Dose Rate
Volume to be Infused
Volume Infused
Time remaining
ADRENALINE
RATE
VOLUME
VTBI
25
.0
ml/h
µg/kg/24h
16.
7
1h 48m 00s
VTBI
VOLUME
ml
ml
45
.0
50
.0
WiFi
Infusion Status /
Drug Name /
Profile Name /
Primary or Secondary
(Only if secondary is
enabled in the data set)
Infusion Rate
Softkeys
Softkey Identifiers
Clear Volume Infused Set VTBI Option
c)
ON HOLD
RATE
SET RATE WITH
ON HOLD
RATE
RATE TOO LOW
ON HOLD
RATE
RATE TOO HIGH
If programmed rate is greater than
the
INFUSION RATE MAX
in drug
protocol, message
c)
will be shown .