BD Alaris™ neXus GP Volumetric Pump
Configuration
BDDF00558 Issue 5
52/72
BD Alaris™ neXus Editor Software Profile Configuration
The following options are only configurable in the data set via the BD Alaris™ neXus Editor Software (PC application) . See
BD Alaris™ neXus Editor
Directions For Use
for further details .
Data Set Configuration Settings
Profile Filtering
Controls whether the User is able to filter which profiles will be available on the Pump .
Unit Display
The text used to display Units, either mU, U and kU or mUnit, Unit and kUnit .
General Pump Configuration Settings
AC Fail Warning
Controls whether, when main power has been disconnected, a warning is generated to
inform the User that the Pump is operating solely on battery power .
Audio Volume
Controls the audio volume used for alarms and warnings .
Audio Volume Adjustable
Controls whether the User is able to adjust the audio volume setting .
Auto Night Mode
Controls whether, between defined times, the Pump makes changes consistent with
usage at night (e .g . dimmed backlight) .
Auto save
Controls whether, when the Pump is powered down, the active infusion settings are
preserved for restoration at the next power-up .
Battery Icon
Controls whether or not the battery icon (which indicates state of charge) is displayed .
Callback Time
The time that is allowed to elapse between the User’s last recorded interaction with
the Pump and generation of a callback alarm .
Drug Override Mode
Always
–
Any changes made to the dose rate that are outside of the Guardrails™ Soft
Alerts will require confirmation before starting infusion .
Smart
– Confirmation of setting will be required on first dose rate set outside of the
Guardrails™ Soft Alert . Any subsequent changes will not require confirmation until
after the dose rate has been confirmed inside the Guardrails™ Soft Alert limits .
Event Log
Controls whether or not it is possible for the User to inspect the contents of the event
log using the Pump’s display and keypad .
Pressure Display
Controls whether or not the bar graph indication of downstream pressure is displayed .
Quiet Mode
Controls whether or not the Pump operates in a mode whereby generation of sound is
minimised .
Rate Lock
Controls whether the Rate Lock feature is available for use .
Rate Titration
Allows the adjustment of the infusion rate while the Pump is infusing, without putting
the Pump on hold .
Standby Mode
Controls whether or not standby mode is available on the Pump .
VTBI Clear Rate
Controls whether or not the Pump forces the User to define a new rate following
completion of delivery of a previous VTBI .
Weight Default
The default patient weight .
Weight Soft Min
The minimum patient weight for weight-based drug dosing calculations before
alerting the User .
Weight Soft Max
The maximum patient weight for weight-based drug dosing calculations before
alerting the User .