BD Alaris™ neXus GP Volumetric Pump
Alarms
BDDF00558 Issue 5
42/72
Display
Infusion Status
Cause(s)
Action
BATTERY EMPTY
Infusion stopped
The internal battery is
exhausted . The Pump will
automatically switch off in
the immediate future .
• Connect to power supply immediately
or switch Pump off .
SAFETY CLAMP
Pump on hold
Safety clamp broken or
missing .
• Clamp infusion set using roller clamp .
• Replace infusion set .
• Investigate and correct set loading .
SET MISLOAD
Pump on hold
Set loaded incorrectly .
• Clamp infusion set using roller clamp .
• Investigate and correct set loading .
FLOW SENSOR
DISCONNECT
Infusion stopped
Flow sensor unplugged
during infusion .
• Check / replace flow sensor or set VTBI .
WRONG SET
Pump on hold
Safety clamp not detected .
• Clamp infusion set using roller clamp .
• Check set and close door .
• Replace infusion set (If necessary) .
DOOR CLOSE
INCOMPLETE
Pump on hold
Safety clamp in non-
occluded position with door
open or obstructed .
• Clamp infusion set using roller clamp .
• Investigate and correct set loading .
• Close door .
DO NOT USE
Pump on hold /
infusion stopped
Internal error has occurred .
• Remove Pump from use .
LEVER OPEN
Infusion stopped
Door lever is open
• Check door lever .
• Check lever hooks .
• Check lever is not obstructed, if so, free
obstruction .
VTBI DONE
Infusion stopped
Intended VTBI completed .
• Set new VTBI or clear VTBI .
SET CLOCK
Pump on hold
Date / time not set .
• Qualified Service Personnel must set
date / time .
• Press
CANCEL
softkey to continue .
SET SERIAL
NUMBER
Pump on hold
Serial number not set .
• Contact Qualified Service Personnel to
set the serial number .