2
HTS Customer Care Kits User’s Guide
Kit Configurations
The HTS Customer Care Kit is designed to custom fit the HTS configuration for
each cytometer or group of cytometers. Therefore, order the correct kit number
by using the following table.
When to Use the Kit
The parts in the HTS Customer Care Kit can be used to resolve common
problems. The following table is lists of symptoms, possible causes, and solutions
that are described in this user’s guide.
Part number
Description
Cytometer
644787
HTS Kit Parts Replacement
BD FACSCanto
™
BD FACSCanto
™
II
644788
HTS Kit Parts Replacement
BD FACSCalibur
™
BD
™
LSR II (including special order
BD LSR II)
Special order BD LSRFortessa
™
Symptom
Possible Causes
Solution
See...
High carryover
Wash station connections
are worn
Replace wash station
components
Chapter 3
Secondary
pump (Pump-2)
error
Clog in sample coupler
tubing, or in pump
syringe, or syringe needs
replacement
Replace sample coupler
tubing or pump
syringes
Chapter 5
Chapter 6
Primary pump
(Pump-1) error
Clog in probe, or tubing
assembly from probe to
the primary pump, or in
pump syringe, or syringe
needs replacement
Replace probe, tubing
assembly components,
or pump syringes
Chapter 4
Chapter 6
Содержание 644787
Страница 3: ...History Revision Date Change Made 23 11165 00 Rev A 1 2010 Initial release ...
Страница 4: ......
Страница 14: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 19: ...Chapter 2 HTS Removal and Reinstallation 13 Captive screw HTS positioned on its support bracket ...
Страница 20: ...14 HTS Customer Care Kits User s Guide Figure 2 1 Electrical connection to the HTS Power plug Serial cable ...
Страница 28: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 32: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 42: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 46: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 50: ...THIS PAGE INTENTIONALLY LEFT BLANK ...