The RUN MODE allows the user to run the test configuration. The screens that are
displayed in the RUN MODE are determined by the Auto Sequence selected on the AUTO
SEQUENCE SCREEN and its configuration as defined with the PC program.
Running an Auto Sequence will provide a consistent, guided procedure for testing
equipment. This is a semi-automated process that will provide immediate feedback to the
user if the DUT passes or fails individual tests. A programmable timer is provided to
automatically progress through the test when a given test passes. This timer is set in the
Auto Sequence Timer parameter in the SYSTEM SETUP SCREEN.
NOTE: If any particular test option is disabled using the PC Program, it will not be shown in
the RUN MODE.
NOTE: Some tests, like Performance Waveforms, do not have quantitative analyses and
therefore require the user to manually progress through the test.
The following sample screen shows the common elements present during the RUN MODE:
RUN MODE
Selected
Auto Sequence
Current
Test Step
Test Results
Exit Key
for exiting the
RUN MODE
NOTE: All Data
will be lost when
this key is used.
Instructions
for Settings and
Actions
Test Results
Passed
Failed
Last Step Key
for returning to
the last step
Next Step Key
for advancing
to the next step
Special Action Key
Capture Readings,
Get New Readings,
Start
Print Test Results Key
for sending all results
to the printer
Содержание DA-2006
Страница 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Страница 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Страница 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Страница 21: ...DA 2006 14 This page intentionally left blank ...
Страница 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Страница 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Страница 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Страница 83: ...DA 2006 Series 76 This page intentionally left blank ...
Страница 85: ...DA 2006 Series 78 This page intentionally left blank ...
Страница 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Страница 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Страница 115: ...DA 2006 Series 108 This page intentionally left blank ...
Страница 122: ...115 NOTES ...
Страница 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Страница 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Страница 139: ...15 NOTES ...
Страница 160: ...D5 ...
Страница 162: ...D8 ...
Страница 165: ...D12 ...
Страница 169: ...D17 ...
Страница 170: ...D18 ...
Страница 171: ...NOTES ...