B 2
OVERVIEW
The DA2006P energy pulse function is based on converting the output from a defibrillator to
a digital value. The defibrillator is discharged into a fixed 50 ohms non-inductive load. The
resulting voltage drop across this resistor is divided by a series of resistors and converted
from an analog voltage to a digital count by the analog-to-digital (A/D) converter. This value
is then processed by the microprocessor and displayed on the LCD. To properly evaluate
the defibrillator pulse, the A/D converter samples 10,000 times per second.
The pace pulse is also analyzed by an A/D converter. The load resistance for the pace
function can be the same load resistance for the defib pulse or a separate set of internal
load resistors can be used. This allows the user to be able to analyze the pace pulse at
many different loads.
ECG waveforms, along with performance waveforms are generated by an internal digital to
analog converter and are available on the ten patient lead connectors.
USER INTERFACE
The user interfaces with the analyzer via the 240x64 pixel Backlit LCD display and the
keys. The LCD display is a full graphical display and is continually updated to inform the
user of the current operating conditions. The seven keys allows the user the ability to setup
and operate the analyzer with a minimal number of keystrokes. The keys labeled F1
through F5 change function based on the current operating mode of the DA2006P. The
lower section of the display is used to identify the function of these keys at any given time.
Содержание DA-2006
Страница 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Страница 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Страница 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Страница 21: ...DA 2006 14 This page intentionally left blank ...
Страница 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Страница 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Страница 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Страница 83: ...DA 2006 Series 76 This page intentionally left blank ...
Страница 85: ...DA 2006 Series 78 This page intentionally left blank ...
Страница 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Страница 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Страница 115: ...DA 2006 Series 108 This page intentionally left blank ...
Страница 122: ...115 NOTES ...
Страница 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Страница 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Страница 139: ...15 NOTES ...
Страница 160: ...D5 ...
Страница 162: ...D8 ...
Страница 165: ...D12 ...
Страница 169: ...D17 ...
Страница 170: ...D18 ...
Страница 171: ...NOTES ...