Defibrillator Analyzer
26
The TEST RESULTS section of the MAIN SCREEN displays the results of the last pulse. It
will continue to be displayed until the power is turned off, another test is run or the range is
changed.
NOTE: The unit automatically starts a test when it sees a voltage greater than 100 Volts on
the Defibrillator Plates (20 Volts in Low Range).
NOTE: Test results are immediately sent to the printer port as soon as the data is
available.
ENERGY
This line displays the total energy of the last pulse.
PEAK V
This line displays the peak voltage of the last pulse.
PEAK I
This line displays the peak current of the last pulse.
TEST RESULTS
Содержание DA-2006
Страница 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Страница 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Страница 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Страница 21: ...DA 2006 14 This page intentionally left blank ...
Страница 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Страница 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Страница 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Страница 83: ...DA 2006 Series 76 This page intentionally left blank ...
Страница 85: ...DA 2006 Series 78 This page intentionally left blank ...
Страница 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Страница 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Страница 115: ...DA 2006 Series 108 This page intentionally left blank ...
Страница 122: ...115 NOTES ...
Страница 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Страница 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Страница 139: ...15 NOTES ...
Страница 160: ...D5 ...
Страница 162: ...D8 ...
Страница 165: ...D12 ...
Страница 169: ...D17 ...
Страница 170: ...D18 ...
Страница 171: ...NOTES ...