Alarms
Manual 41018v0800
SIGMA Spectrum Infusion System
Revision B
Operator’s Manual
PRIMARY INFUSION
COMPLETE
Displays when the Primary Infusion VTBI has counted down to zero.
■
The Pump is running at a Keep Vein Open rate (KVO rate) or the actual infusion
rate, whichever is lower.
■
The Drug Library default KVO rate is set at 1 mL/hr in the Drug Library.
RELOAD SET
Displays when the Pump door is closed and the IV tube has not been loaded in the
correct order or in all the load points within the tubing channel.
■
Unload and reload the IV set. See “Loading an IV Set” on page 37 and
“Unloading an IV Set” on page 44.
REMOVE CLAMP
FROM PRIMARY
Displays at the completion of a secondary infusion if
no
is selected during the
Secondary Check Flow tutorial and
yes
is selected in response to Clamp Applied?
during tutorial. See “Secondary Infusion” on page 58.
At the end of a secondary infusion:
1.
Press the
review Primary
soft key.
2.
Remove applied clamp from primary line.
SECONDARY
INFUSION COMPLETE
Displays when the programmed secondary VTBI has counted down to zero and
secondary callback is applied by either the Drug Library or has been selected. The
Pump runs at KVO rate.
If Secondary Callback is unavailable or not selected, the Pump transitions to the
delivery of the primary program.
SLIDE CLAMP
DETECTED
Displays when the Pump detects a slide clamp in the keyhole.
1.
Remove slide clamp from the keyhole.
2.
Press RUN.
3.
If slide clamp not present in keyhole and the alarm occurs, inspect for debris.
4.
If the alarm re-occurs, send the Pump for service to qualified personnel at your
facility.