Appendix G: Default Settings
161
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
BASIC Configurations
A BASIC infusion provides non-DERS based infusion programming.
DERS limits do not exist when using BASIC mode. BASIC mode allows the user to manually
specify an mL/hr rate, dose mode, dose rate and volume to be infused, among other parameters. It
does possess the following safety features: Check Flow at Run, Secondary Error Prevention and
Single Step Rate Change of 101%.
CAUTION
BASIC Programming Use.
BASIC programming should only be used when the desired drug or
concentration is not available in the facility’s Drug Library.
Table 3. BASIC Fixed Settings.
BASIC Configuration Settings
Fixed Settings
Allow Bolus
Yes
Allow Loading Dose
No
Bolus Amount Units
mL
Bolus Time Units
Minutes and Seconds (00:00)
Completion Alarm
Not Available
Delayed Run
Not Available
KVO Rate
1 mL/hr
Secondary Callback
Never
Single Step Rate Change
101%
Table 4. BASIC Settings Configurable on the Pump.
BASIC Configuration Settings
Default Settings
Audio Alarm
High, Medium, Low
Bag Near Empty Alarm
On, Off
Delivery Bag
Primary Bag, Secondary Bag
Dose Mode
Clinician can select