Programming the Pump
99
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
NOTE:
If the standby period is set to infinite, the time value in the display will be replaced
with a dashed line.
3. While the Pump is in standby, press
RUN
at any time to begin the infusion.
NOTE:
Pressing any other key or opening the Pump door cancels standby mode.
4. To start the infusion, ensure that the slide clamp and roller clamp are open and press
RUN
.
Once
RUN
is pressed, the infusion will start and the Primary Check Flow screen is displayed.
Check the flow and confirm:
All clamps are open
There are no kinks or collapses in the tubing outside of the Pump
Drops are flowing in the drip chamber. Note, at very low rates, it may take several minutes
to see drops.
Vents are open (if applicable).
If drops are falling in the drip chamber, press the
yes
soft key.
If no drops are falling, press the
no
soft key and follow the screen prompts.
To place a Pump in standby after an infusion has started:
1. Press
STOP
.
2. Press the
review/program
soft key.
3. Press the
hold
soft key.
Figure 126. In STANDBY Message.