Programming the Pump
91
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
Secondary Line Flush
NOTE:
Important
. For the drug products mentioned, refer to the manufacturers’ package insert
for full prescribing information which may include boxed warnings.
At the completion of the secondary infusion a dialog box will be displayed to enable the line flush
only if:
■
Yes is pressed to request the secondary callback on the dialog box or,
■
Secondary Callback has been configured as required in the Drug Library
NOTE:
In the Optional setting, if no is selected to a callback, Secondary Line Flush will not
be available.
The Secondary Line Flush feature allows delivery of any residual fluid volume remaining in the IV
container of a secondary Amount/Time infusion.
■
Line Flush is available for secondary infusions when the Secondary Callback feature is set to
Required or Optional in the Drug Library.
■
Line Flush infuses at the same rate as the last programmed infusion rate.
■
Line Flush feature may be repeated up to three times.
■
Line Flush volume can be set to between 1 mL and 100 mL.
■
Line Flush is only available for Amount/Time infusions.
■
Line Flush volume is set in the Drug Library and can be changed at the Pump.
To flush the secondary line:
1. At the completion of the secondary infusion, the Secondary Complete screen will display if the
secondary callback has been enabled or configured for this drug in the Drug Library. See
Figure 108. Also see Figure 109.
2. Press the
yes
soft key to deliver the residual volume (line flush).
Press the
no
soft key to decline the flush.
Figure 108. Secondary Complete Alarm.
Figure 109. Secondary Complete: Line
Flush Dialog Box.