58
Programming the Pump
Manual 41018v0800
SIGMA Spectrum Infusion System
Revision B
Operator’s Manual
2. Press the
review
or
program pri/sec
soft key to display the Setup screen.
3. Press the
clear total
soft key to clear the total given.
To change the rate while the Pump is stopped:
1. Press
STOP
.
2. Press the
review
or
program pri/sec
soft key to display the Setup screen.
3. Press the arrow soft keys to select the desired value.
4. Enter the new rate value.
5. Press
OK
to confirm.
6. Press
RUN
to resume the infusion.
Secondary Infusion
To program a secondary infusion:
1. Prepare the IV sets and container for a secondary infusion using the steps in Secondary
2. Follow the steps for programming the primary bag infusion as described in either Dose Error
Reduction System (DERS) Programming on page 50 or BASIC Mode on page 94.
NOTE:
The SIGMA Spectrum Infusion Pump does not deliver a primary and secondary
infusion at the same time.
3. Press the arrow soft keys to select a primary bag or a secondary bag and press
OK
. See
NOTE:
The bag selection prompt does not appear if the selected drug has been specifically
assigned in the Drug Library as:
Figure 40. Bag Selection Screen.