6-12 Operator’s Manual
110017276EN Rev. B
6 Troubleshooting
ID Type
Text Message
Suggested Action(s)
USM08
Info
The ultrasound handpiece has failed the
tuning process.
SG52 Ensure ultrasound needle is properly
tightened. Select ‘Prime / Tune’ or ‘Tune
Only’ again to repeat the handpiece tuning
process.
SG53 Unplug the handpiece and try another
ultrasound handpiece.
SG04 Call your product service
representative if this problem persists.
USM09
Info
The ultrasound handpiece may have
failed or may be failing.
SG54 Confirm proper ultrasound handpiece
operation in a test chamber.
SG53 Unplug the handpiece and try another
ultrasound handpiece.
SG04 Call your product service
representative if this problem persists.
USM10
Warning
Coagulation circuit may have failed,
potential coagulation output over-voltage
condition.
SG55 Confirm proper operation of the
coagulation handpiece in a beaker of BSS;
activate the function and ensure the error
message does appear again. If similar
message appears, stop using the function
and call B&L service support.
SG56 Unplug the handpiece and try another
bipolar coagulation cord.
SG04 Call your product service
representative if this problem persists.
USM11
Warning
Coagulation circuit may have failed,
potential uncommanded or incorrect
coagulation output.
SG55 Confirm proper operation of the
coagulation handpiece in a beaker of BSS;
activate the function and ensure the error
message does appear again. If similar
message appears, stop using the function
and call B&L service support.
SG56 Unplug the handpiece and try another
bipolar coagulation cord.
SG04 Call your product service
representative if this problem persists.
USM15
Info
Unable to read ultrasound handpiece
data. Attempt to tune the handpiece to
confirm proper operation.
SG53 Unplug the handpiece and try another
ultrasound handpiece.
SG04 Call your product service
representative if this problem persists.
Содержание STELLARIS PC
Страница 1: ...Operator s Manual ...
Страница 24: ...1 14 Operator s Manual 110017276EN Rev B 1 Getting Started Figure 1 9 Posterior Domain Case Menu Screen ...
Страница 34: ...1 24 Operator s Manual 110017276EN Rev B 1 Getting Started Figure 1 14 Anterior Surgical Screen ...
Страница 36: ...1 26 Operator s Manual 110017276EN Rev B 1 Getting Started Figure 1 16 Anterior End of Case Screen ...
Страница 37: ...Operator s Manual 1 27 110017276EN Rev B 1 Getting Started Figure 1 17 Posterior End of Case Screen ...
Страница 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Страница 88: ...2 10 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 14 Vitrectomy Cut More Settings Screen tab ...
Страница 90: ...2 12 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 16 Phaco More Settings Screen tab ...
Страница 93: ...Operator s Manual 2 15 110017276EN Rev B 2 User Interface Figure 2 19 Extrude More Settings Screen tab ...
Страница 94: ...2 16 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 20 I A More Settings Screen tab ...
Страница 95: ...Operator s Manual 2 17 110017276EN Rev B 2 User Interface Figure 2 21 Irrigation More Settings Screen tab ...
Страница 99: ...Operator s Manual 2 21 110017276EN Rev B 2 User Interface Figure 2 25 Phaco More Settings Screen Tab ...
Страница 116: ...2 38 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 41 Small Video Combined Surgery Screen ...
Страница 120: ...2 42 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 45 Sample Drop Down Option List ...
Страница 128: ...2 50 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 55 Anterior Vitrectomy More Settings Screen Cut tab ...
Страница 129: ...Operator s Manual 2 51 110017276EN Rev B 2 User Interface Figure 2 56 Phaco More Settings Screen Ultrasound tab ...
Страница 130: ...2 52 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 57 I A More Settings Screen Infusion tab ...
Страница 131: ...Operator s Manual 2 53 110017276EN Rev B 2 User Interface Figure 2 58 Irrigation More Settings Screen Infusion tab ...
Страница 148: ...2 70 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 74 Large Video Anterior Surgery Screen ...
Страница 149: ...Operator s Manual 2 71 110017276EN Rev B 2 User Interface Figure 2 75 Small Video Anterior Surgery Screen ...
Страница 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Страница 207: ...Operator s Manual 4 17 110017276EN Rev B 4 Detailed Reference Figure 4 11 Lamp Housing Exterior ...
Страница 234: ...4 44 Operator s Manual 110017276EN Rev B 4 Detailed Reference Figure 4 26 More Settings Screen Infusion Tab Anterior ...
Страница 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Страница 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Страница 300: ...8 4 Operator s Manual 110017276EN Rev B 8 Specifications Cable Length BL4394 Foot Control Power Cable 144 ...
Страница 316: ...8 20 Operator s Manual 110017276EN Rev B 8 Specifications Xenon Lamp Label Xenon Mercury Lamp Label ...
Страница 329: ...Preface ...
Страница 330: ...Table of Contents ...
Страница 331: ...Getting Started ...
Страница 332: ...User Interface ...
Страница 333: ...Customizing Your System ...
Страница 334: ...Detailed Reference ...
Страница 335: ...Cleaning and Sterilization ...
Страница 336: ...Troubleshooting ...
Страница 337: ...Service and Warranty ...
Страница 338: ...Specifications ...
Страница 339: ...Glossary ...
Страница 340: ...Index ...
Страница 341: ...Preface ...
Страница 342: ...Table of Contents ...
Страница 343: ...Getting Started ...
Страница 344: ...User Interface ...
Страница 345: ...Customizing Your System ...
Страница 346: ...Detailed Reference ...
Страница 347: ...Cleaning and Sterilization ...
Страница 348: ...Troubleshooting ...
Страница 349: ...Service and Warranty ...
Страница 350: ...Specifications ...
Страница 351: ...Glossary ...
Страница 352: ...Index ...