from any ultrasonic handpiece may be the result of ultrasonic energy causing micro abrasion of
the ultrasonic tip.
Fragmentation modes
The
Stellaris® PC
Vision Enhancement System provides two Fragmentation modes with adjustable ultrasound
power and vacuum control. Refer to the table on page 3-33 for information on the available vacuum range and
options on ultrasound modulations. Detailed description of Ultrasound modulations such as Pulse, Burst and
Duty Cycle are provided in the Ultrasound Functions section of this chapter.
Linear Frag
Linear Frag submode provides linear vacuum in footpedal region 1 and fixed maximum vacuum and linear
ultrasound in food pedal region 2. The outward yaw in any footpedal region controls ultrasound On/Off.
Dual/Yaw US Frag
This submode provides linear vacuum control on footpedal pitch, and linear ultrasound on outward yaw in
footpedal region 1.
Fragmentation Setup and Use:
Note:
Prior to setting up Fragmentation, the
Stellaris® PC
Vision Enhancement System has to be
setup for Posterior Vitrectomy or Combined surgery.
Note:
Specific instructions for cleaning and sterilization included with a handpiece or accessory take
precedence over these instructions.
WARNING:
A loose needle may lead to improper tuning and could cause shedding of metal fragments into
the eye, which can result in serious permanent patient injury.
1.
Setup the system for posterior vitrectomy ( page 4-7 ) or combined surgery ( page 4-50 ).
2.
Ensure the screen is displaying the Prime/Tune Setup screen.
3.
Connect the handpiece electrical connector to the connector on the machine. (Scrub nurse or Circulating
nurse)
4.
Thread and firmly secure the single use fragmentation needle onto the fragmentation handpiece using a
needle wrench. (Scrub nurse)
Note:
Pack number BL5275 would be appropriate for use.
5.
Connect the clear aspiration line to the fragmentation handpiece. (Scrub nurse)
Содержание STELLARIS PC
Страница 1: ...Operator s Manual ...
Страница 24: ...1 14 Operator s Manual 110017276EN Rev B 1 Getting Started Figure 1 9 Posterior Domain Case Menu Screen ...
Страница 34: ...1 24 Operator s Manual 110017276EN Rev B 1 Getting Started Figure 1 14 Anterior Surgical Screen ...
Страница 36: ...1 26 Operator s Manual 110017276EN Rev B 1 Getting Started Figure 1 16 Anterior End of Case Screen ...
Страница 37: ...Operator s Manual 1 27 110017276EN Rev B 1 Getting Started Figure 1 17 Posterior End of Case Screen ...
Страница 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Страница 88: ...2 10 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 14 Vitrectomy Cut More Settings Screen tab ...
Страница 90: ...2 12 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 16 Phaco More Settings Screen tab ...
Страница 93: ...Operator s Manual 2 15 110017276EN Rev B 2 User Interface Figure 2 19 Extrude More Settings Screen tab ...
Страница 94: ...2 16 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 20 I A More Settings Screen tab ...
Страница 95: ...Operator s Manual 2 17 110017276EN Rev B 2 User Interface Figure 2 21 Irrigation More Settings Screen tab ...
Страница 99: ...Operator s Manual 2 21 110017276EN Rev B 2 User Interface Figure 2 25 Phaco More Settings Screen Tab ...
Страница 116: ...2 38 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 41 Small Video Combined Surgery Screen ...
Страница 120: ...2 42 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 45 Sample Drop Down Option List ...
Страница 128: ...2 50 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 55 Anterior Vitrectomy More Settings Screen Cut tab ...
Страница 129: ...Operator s Manual 2 51 110017276EN Rev B 2 User Interface Figure 2 56 Phaco More Settings Screen Ultrasound tab ...
Страница 130: ...2 52 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 57 I A More Settings Screen Infusion tab ...
Страница 131: ...Operator s Manual 2 53 110017276EN Rev B 2 User Interface Figure 2 58 Irrigation More Settings Screen Infusion tab ...
Страница 148: ...2 70 Operator s Manual 110017276EN Rev B 2 User Interface Figure 2 74 Large Video Anterior Surgery Screen ...
Страница 149: ...Operator s Manual 2 71 110017276EN Rev B 2 User Interface Figure 2 75 Small Video Anterior Surgery Screen ...
Страница 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Страница 207: ...Operator s Manual 4 17 110017276EN Rev B 4 Detailed Reference Figure 4 11 Lamp Housing Exterior ...
Страница 234: ...4 44 Operator s Manual 110017276EN Rev B 4 Detailed Reference Figure 4 26 More Settings Screen Infusion Tab Anterior ...
Страница 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Страница 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Страница 300: ...8 4 Operator s Manual 110017276EN Rev B 8 Specifications Cable Length BL4394 Foot Control Power Cable 144 ...
Страница 316: ...8 20 Operator s Manual 110017276EN Rev B 8 Specifications Xenon Lamp Label Xenon Mercury Lamp Label ...
Страница 329: ...Preface ...
Страница 330: ...Table of Contents ...
Страница 331: ...Getting Started ...
Страница 332: ...User Interface ...
Страница 333: ...Customizing Your System ...
Страница 334: ...Detailed Reference ...
Страница 335: ...Cleaning and Sterilization ...
Страница 336: ...Troubleshooting ...
Страница 337: ...Service and Warranty ...
Страница 338: ...Specifications ...
Страница 339: ...Glossary ...
Страница 340: ...Index ...
Страница 341: ...Preface ...
Страница 342: ...Table of Contents ...
Страница 343: ...Getting Started ...
Страница 344: ...User Interface ...
Страница 345: ...Customizing Your System ...
Страница 346: ...Detailed Reference ...
Страница 347: ...Cleaning and Sterilization ...
Страница 348: ...Troubleshooting ...
Страница 349: ...Service and Warranty ...
Страница 350: ...Specifications ...
Страница 351: ...Glossary ...
Страница 352: ...Index ...