[007] SIGNAL ERROR
The measurement is heavily influ-
enced by external interference or the
connection cable is damaged.
n
Check the cable connections.
n
Check ground connections.
n
Eliminate possible sources of
interference.
[008] PIPE EMPTY
The measuring pipe is empty or the
sensor has not been properly cali-
brated.
n
Check the fill level of the meas-
uring tube.
n
Check the empty calibration of
the sensor.
[009] FLOW>MAX+
The flow rate is higher than the set
maximum positive threshold value.
n
Check process conditions.
n
Check the set maximum
threshold for the positive flow rate
in the
Scales
menu.
[010] FLOW>MAX-
The flow rate is higher than the set
negative maximum threshold.
n
Check process conditions.
n
Check the set maximum
threshold for the negative flow
rate in the
Scales
menu.
[011] FLOW<MIN+
The flow rate is lower than the set
minimum positive threshold.
n
Check process conditions.
n
Check the set minimum positive
flow rate threshold in the
Scales
menu.
[012] FLOW<MIN+
The flow rate is lower than the set
minimum negative threshold.
n
Check process conditions.
n
Check the set minimum negative
flow rate threshold in the
Scales
menu.
[013] FLOW>FULL SCALE
+
The flow rate is higher than the set
positive scale end value.
n
Check process conditions.
n
Check the set maximum value for
the positive flow rate in the
Scales
menu.
[014] FLOW>FULL SCALE- The flow rate is lower than the set
negative scale end value.
n
Check process conditions.
n
Check the maximum negative
flow rate set in the
Scales
menu.
[015] PULSE1>RANGE
The pulse generator of the Output 1
of the sensor can not generate the
sufficient number of pulses.
n
Increase set value for volume unit
in the
Units
menu.
or
n
Reduce the set interval for pulse
generation in the
Scales
menu.
[016] PULSE2> RANGE
The pulse generator of the Output 2
of the sensor can not generate the
sufficient number of pulses.
n
Increase set value for volume unit
in the
Units
menu.
or
n
Reduce the set interval for pulse
generation in the
Scales
menu
.
[017] CALIBR.ERROR
Calibration error
Contact Service.