Description
Password level /
BCP command
Calibration
External calibration command
If the function is activated, zero calibration can
be performed by an external signal.
n
If the voltage pulse of the external signal is
shorter than 1 s, the instrument performs a
calibration cycle to compensate for any
potential thermal variations.
n
If the voltage pulse of the external signal is
longer than 1 s, the meter performs a zero
calibration measurement.
For zero calibration, make sure that the meas-
uring pipe is completely filled with liquid and the
liquid is completely still. Even the smallest
movements of liquid can cause considerable
measuring errors.
L 3
CALIE
Range change
Change of the measuring range
If the function is activated, the measuring range
can be changed by an external signal.
This function cannot be activated if the
Autorange
function is activated.
L 3
SRCIE
Operation
Menu navigation > Inputs menu
05.03.2020
Flow sensor PF55S