background image

BARTINGTON INSTRUMENTS

 

Page 7 of 9 

OM2497/4

6. Troubleshooting

The sensor is unlikely to suffer any defects in normal use: no internal components are 
serviceable. The most likely causes of failure, and their solutions, are detailed in the following 
table. 

In the event of any apparent malfunction beyond those described in the table below, please email 

[email protected]

, or telephone the Bartington Instruments service team on +44 (0)1993 

706565. 

Causes of Failure

Solution

Power supply failure

Check the power, as detailed in 

Pre-

Installation Tests

Cables

Test the cables, as detailed in 

Pre-

Installation Tests

.

Power input

If no fault can be found in the power supply 
or cables, ensure the cable length is not 
too long, causing excessive voltage drop 
between the power supply and sensor. 
Refer to the specifications defined in the 
product brochure.

Sensor damage

Physical damage to the Mag670 cannot 
be repaired. Replace with a new unit. See 

Disposal

.

7. Care and Maintenance

No servicing is normally required. No repair is possible. For further details refer to 

Troubleshooting

.

7.1. Cleaning Mag670 Sensors

Caution: 

To clean an unenclosed variant of the Mag670, use an air duster to blow out 

dust and debris if required. Use appropriate personal protective equipment, such as eye 
protection, when carrying out this task.

Caution:

 Avoid using chemicals, such as solvents, when cleaning the Mag670. Use a damp 

cloth to remove grime from enclosures (packaged version only).

Содержание Mag670 Series

Страница 1: ...www bartington com INNOVATION IN MAGNETICS Operation Manual for Mag670 Single Axis Magnetic Field Sensor OM 2497 5...

Страница 2: ...nment Recommendations 4 4 2 Cable Recommendations 4 4 3 Pre Installation Tests 5 4 4 Mounting Recommendations 5 4 5 Post Installation Testing 6 5 Using the Mag670 6 5 1 Magnetic Hysteresis 6 5 2 Envir...

Страница 3: ...hich bodily injury or damage to your instrument or both could result if the caution is ignored Indicates a situation in which ESD protection should be used Identifies items that must be disposed of sa...

Страница 4: ...Site the sensor several tens of metres from very large ferromagnetic objects that could create fields exceeding the measuring range of the sensor Note Avoid siting the sensor near any ferromagnetic ob...

Страница 5: ...a Test the cables end to end at the connectors to ensure that the correct pins have been allocated to the conductors and that there are no open or high resistance circuits Note Cable resistance will v...

Страница 6: ...the sensor should remain stable to within the quoted noise limits Refer to the product brochure for the expected values 5 Using the Mag670 5 1 Magnetic Hysteresis Caution Subjecting the sensor to fiel...

Страница 7: ...o fault can be found in the power supply or cables ensure the cable length is not too long causing excessive voltage drop between the power supply and sensor Refer to the specifications defined in the...

Страница 8: ...on on using this chemical safely 7 2 Calibration Return the Mag670 to Bartington Instruments for calibration at the recommended intervals Refer to the Calibration Certificate for further details 8 End...

Страница 9: ...OX28 4GE England The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Austral...

Отзывы: