Note: 

The length of the cable is limited by the voltage drop in the power supply lines 

and the capacitance between the cores. For this reason the cable should be limited to a 
maximum length of 600 metres.

Bartington Instruments can supply cables for connection of the sensor to the PSU1, Magmeter, 
SCU1 and Spectramag-6 data acquisition and power supply units.

4.3. Pre-Installation Tests

Prior to the installation of the system, the sensor, cable and power supply must be fully tested to 
ensure correct function as follows:

Caution:

 Take care to avoid bending or otherwise damaging the contacts whilst conducting 

the tests.

1.  Test the cables for continuity (using an electrical continuity tester or ohmmeter).

a.  Test the cables end to end at the connectors, to ensure that the correct pins have 

been allocated to the conductors and that there are no open or high resistance 
circuits.

Note: 

Cable resistance will vary: refer to the product brochure for the expected values.

b.  Test the cables at the connectors to ensure that there are no short circuits between 

the conductors.

2.  Check the power supply output voltage using a voltmeter. Refer to the product brochure 

for the expected values.

3.  Connect the sensor to the cable connector.

4.  Connect the power supply to the other cable connector.

4.4. Mounting Recommendations

Each sensor in both packaged and unpackaged versions has a set of mounting holes to allow 
attachment to a stable base or fixture. Refer to drawings DR2545 and DR2945 for further 
information.

The sensor can also mounted on the Bartington Instruments Mag-TA Universal Tripod Adaptor. 
See the product brochure for instructions on how to do this.

Содержание Mag670 Series

Страница 1: ...www bartington com INNOVATION IN MAGNETICS Operation Manual for Mag670 Single Axis Magnetic Field Sensor OM 2497 5...

Страница 2: ...nment Recommendations 4 4 2 Cable Recommendations 4 4 3 Pre Installation Tests 5 4 4 Mounting Recommendations 5 4 5 Post Installation Testing 6 5 Using the Mag670 6 5 1 Magnetic Hysteresis 6 5 2 Envir...

Страница 3: ...hich bodily injury or damage to your instrument or both could result if the caution is ignored Indicates a situation in which ESD protection should be used Identifies items that must be disposed of sa...

Страница 4: ...Site the sensor several tens of metres from very large ferromagnetic objects that could create fields exceeding the measuring range of the sensor Note Avoid siting the sensor near any ferromagnetic ob...

Страница 5: ...a Test the cables end to end at the connectors to ensure that the correct pins have been allocated to the conductors and that there are no open or high resistance circuits Note Cable resistance will v...

Страница 6: ...the sensor should remain stable to within the quoted noise limits Refer to the product brochure for the expected values 5 Using the Mag670 5 1 Magnetic Hysteresis Caution Subjecting the sensor to fiel...

Страница 7: ...o fault can be found in the power supply or cables ensure the cable length is not too long causing excessive voltage drop between the power supply and sensor Refer to the specifications defined in the...

Страница 8: ...on on using this chemical safely 7 2 Calibration Return the Mag670 to Bartington Instruments for calibration at the recommended intervals Refer to the Calibration Certificate for further details 8 End...

Страница 9: ...OX28 4GE England The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Austral...

Отзывы: