background image

BARTINGTON INSTRUMENTS

 

Page 6 of 9 

OM2497/4

4.5. Post Installation Testing

1.  Switch on the power supply and wait until the sensor has stabilised. Refer to the 

product brochure for warm-up times.

2.  Confirm no magnetic objects are moving in the vicinity.

3.  Check the output:

a.  connect a voltmeter between output and GND;

b.  whilst monitoring the voltmeter readings, align the sensor with the terrestrial field 

until the maximum voltage value is determined;

c.  confirm that the measured reading approaches the local geomagnetic field value.

Note:

 Geomagnetic field values can be provided by your local magnetic observatory. A 

margin of error due to local disturbance should be taken into account.

Note:

 The outputs from the sensor should remain stable to within the quoted noise limits. 

Refer to the product brochure for the expected values.

5. Using the Mag670

5.1. Magnetic Hysteresis

Caution:

 Subjecting the sensor to fields in excess of 2 x the nominal range may cause 

inaccuracy in future measurements. Degaussing the sensor can reverse such an effect.

Note: 

The Mag670 is designed to have an extremely low magnetic hysteresis. However, 

Bartington Instruments recommends your sensor is not subjected to magnetic fields 
greater than their stated measuring range for extended periods as this could alter the 
DC offset. If this occurs, the offset will exhibit drift as it returns to its original offset 
specification.

5.2. Environmental Precautions

Refer to the product brochure for maximum environmental electrical and mechanical ratings.

Caution: 

Exceeding the maximum environmental ratings may cause irreparable damage to 

your sensor.

Содержание Mag670 Series

Страница 1: ...www bartington com INNOVATION IN MAGNETICS Operation Manual for Mag670 Single Axis Magnetic Field Sensor OM 2497 5...

Страница 2: ...nment Recommendations 4 4 2 Cable Recommendations 4 4 3 Pre Installation Tests 5 4 4 Mounting Recommendations 5 4 5 Post Installation Testing 6 5 Using the Mag670 6 5 1 Magnetic Hysteresis 6 5 2 Envir...

Страница 3: ...hich bodily injury or damage to your instrument or both could result if the caution is ignored Indicates a situation in which ESD protection should be used Identifies items that must be disposed of sa...

Страница 4: ...Site the sensor several tens of metres from very large ferromagnetic objects that could create fields exceeding the measuring range of the sensor Note Avoid siting the sensor near any ferromagnetic ob...

Страница 5: ...a Test the cables end to end at the connectors to ensure that the correct pins have been allocated to the conductors and that there are no open or high resistance circuits Note Cable resistance will v...

Страница 6: ...the sensor should remain stable to within the quoted noise limits Refer to the product brochure for the expected values 5 Using the Mag670 5 1 Magnetic Hysteresis Caution Subjecting the sensor to fiel...

Страница 7: ...o fault can be found in the power supply or cables ensure the cable length is not too long causing excessive voltage drop between the power supply and sensor Refer to the specifications defined in the...

Страница 8: ...on on using this chemical safely 7 2 Calibration Return the Mag670 to Bartington Instruments for calibration at the recommended intervals Refer to the Calibration Certificate for further details 8 End...

Страница 9: ...OX28 4GE England The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Austral...

Отзывы: