background image

BARTINGTON INSTRUMENTS

 

Page 3 of 9 

OM2497/4

1. About this Manual

This manual provides the information necessary to help customers connect, install and operate, 
the Mag670 magnetic field sensor.

Technical specifications of the products, including power supply requirements and analogue 
output details, can be found in 

DS2497

, whilst outline drawings of both sensor head, electronics 

board and cable can be found on 

DR2545

.

1.1. Symbols Glossary

The following symbols used within this manual call your attention to specific types of 
information:

WARNING: 

Indicates a situation in which serious bodily injury or death could result if the 

warning is ignored.

Caution:

 Indicates a situation in which bodily injury or damage to your instrument, or both, 

could result if the caution is ignored.

Indicates a situation in which ESD protection should be used.

Identifies items that must be disposed of safely to prevent unnecessary damage to the 
environment.

Note:

 

A paragraph in this format provides useful supporting information on how to make 

better use of your purchase.

2. Safe Use

WARNING:

 These products are not qualified for use in explosive atmospheres or life 

support systems. Consult Bartington Instruments for advice.

WARNING: 

Environmental and electrical specifications should not be exceeded.

To prevent irreparable damage, electrostatic discharge (ESD) protection and precautions 
must be used when handling the unpackaged sensor electronics board.

Note: Do not expose to strong magnetic fields while being stored as this can magnetise 
the sensor and affect its offset performance.

Содержание Mag670 Series

Страница 1: ...www bartington com INNOVATION IN MAGNETICS Operation Manual for Mag670 Single Axis Magnetic Field Sensor OM 2497 5...

Страница 2: ...nment Recommendations 4 4 2 Cable Recommendations 4 4 3 Pre Installation Tests 5 4 4 Mounting Recommendations 5 4 5 Post Installation Testing 6 5 Using the Mag670 6 5 1 Magnetic Hysteresis 6 5 2 Envir...

Страница 3: ...hich bodily injury or damage to your instrument or both could result if the caution is ignored Indicates a situation in which ESD protection should be used Identifies items that must be disposed of sa...

Страница 4: ...Site the sensor several tens of metres from very large ferromagnetic objects that could create fields exceeding the measuring range of the sensor Note Avoid siting the sensor near any ferromagnetic ob...

Страница 5: ...a Test the cables end to end at the connectors to ensure that the correct pins have been allocated to the conductors and that there are no open or high resistance circuits Note Cable resistance will v...

Страница 6: ...the sensor should remain stable to within the quoted noise limits Refer to the product brochure for the expected values 5 Using the Mag670 5 1 Magnetic Hysteresis Caution Subjecting the sensor to fiel...

Страница 7: ...o fault can be found in the power supply or cables ensure the cable length is not too long causing excessive voltage drop between the power supply and sensor Refer to the specifications defined in the...

Страница 8: ...on on using this chemical safely 7 2 Calibration Return the Mag670 to Bartington Instruments for calibration at the recommended intervals Refer to the Calibration Certificate for further details 8 End...

Страница 9: ...OX28 4GE England The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Austral...