background image

BARTINGTON INSTRUMENTS

 

Page 4 of 9 

OM2497/4

3. Introduction to the Mag670

The Mag670 is a single-axis fluxgate sensor that provides highly linear magnetic response and 
low hysteresis. Each instrument consists of a single, feedback-stabilised, fluxgate sensing 
element arranged along the length of the enclosure, which is orientated to point away from the 
unit’s connector.

The sensor has no power supply or data processing abilities of its own and must always be 
connected to a power supply to be able to function.

Regulating the power supply internally ensures that the Mag670 is suitable for battery-powered 
operation, over both long and short cables.

Packaged and unpackaged versions are available.

4. Installing the Mag670

4.1. Siting the Sensor (Environment Recommendations)

Note: 

Site the sensor several metres from any magnetic base rock to avoid compromising 

measurements.

Note: 

Site the sensor several tens of metres from very large ferromagnetic objects that 

could create fields exceeding the measuring range of the sensor.

Note: 

Avoid siting the sensor near any ferromagnetic objects that may be subjected to the 

effects of magnetic hysteresis, which would affect the sensor in an unpredictable manner.

Note: 

Conduct a magnetic evaluation of any proposed installation site to establish that 

it is free from magnetic contaminants. Carry out such an evaluation using total field or 
resonance sensors.

4.2. Cable Recommendations

Note: 

The connecting cable to the sensor should be a five-core screened cable. Two cores 

will be used for positive and negative power supply lines, one core for output signal, one 
core for signal common, and one for power supply ground. The screen should be connected 
to supply ground at the supply end only. The capacitance between cores should be less 
than 200pF per metre.

Note: 

Pin out values are shown in drawings DR2545 and DR2945.

Note: 

Leads are susceptible to electromagnetic interference and should be screened 

wherever possible.

Содержание Mag670 Series

Страница 1: ...www bartington com INNOVATION IN MAGNETICS Operation Manual for Mag670 Single Axis Magnetic Field Sensor OM 2497 5...

Страница 2: ...nment Recommendations 4 4 2 Cable Recommendations 4 4 3 Pre Installation Tests 5 4 4 Mounting Recommendations 5 4 5 Post Installation Testing 6 5 Using the Mag670 6 5 1 Magnetic Hysteresis 6 5 2 Envir...

Страница 3: ...hich bodily injury or damage to your instrument or both could result if the caution is ignored Indicates a situation in which ESD protection should be used Identifies items that must be disposed of sa...

Страница 4: ...Site the sensor several tens of metres from very large ferromagnetic objects that could create fields exceeding the measuring range of the sensor Note Avoid siting the sensor near any ferromagnetic ob...

Страница 5: ...a Test the cables end to end at the connectors to ensure that the correct pins have been allocated to the conductors and that there are no open or high resistance circuits Note Cable resistance will v...

Страница 6: ...the sensor should remain stable to within the quoted noise limits Refer to the product brochure for the expected values 5 Using the Mag670 5 1 Magnetic Hysteresis Caution Subjecting the sensor to fiel...

Страница 7: ...o fault can be found in the power supply or cables ensure the cable length is not too long causing excessive voltage drop between the power supply and sensor Refer to the specifications defined in the...

Страница 8: ...on on using this chemical safely 7 2 Calibration Return the Mag670 to Bartington Instruments for calibration at the recommended intervals Refer to the Calibration Certificate for further details 8 End...

Страница 9: ...OX28 4GE England The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Austral...