Bus systems
8-7
HYGROPHIL® F 5673 Operating manual, Software version 1.8.x, 387896MDHEN(14.06.2013)
Table 3 continued
Starting on byte 4 to byte 115 the system mirrors four-byte float values.
Channel 1 Data
4-7
TT
C
8-11
SP
hPa/mbar
12-15
WL
nm
16-19
VOL
%
20-23
PPM
ppm
24-27
DT
C
28-31
FP
C
32-35
MC
mg/m³
36-39
VP
hPa/mbar
Channel 2 Data
40-43
TT
C
44-47
SP
hPa/mbar
48-51
WL
nm
52-55
VOL
%
56-59
PPM
ppm
60-63
DT
C
64-67
FP
C
68-71
MC
mg/m³
72-75
VP
hPa/mbar
Channel 3 Data
76-79
TT
C
80-83
SP
hPa/mbar
84-87
WL
nm
88-91
VOL
%
92-95
PPM
ppm
96-99
DT
C
100-103
FP
C
104-107
MC
mg/m³
108-111
VP
hPa/mbar
HCDT Data
112-115
HCDT
C
116-122
reserved
Содержание HYGROPHIL F 5673
Страница 2: ......
Страница 15: ...System description 1 9 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 28: ...System description 1 22 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 32: ...Safety precautions 2 4 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 43: ...Installation 3 11 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 110: ...Error handling 6 4 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 122: ...Bus systems 8 10 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 128: ......
Страница 129: ...HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...