Programming
5-40
HYGROPHIL® F 5673 Operating manual, Software version 1.8.x, 387896MDHEN(14.06.2013)
● Open the menu “Offset”.
● Choose one of the three possible measurement channels for which you
want to program an offset.
● Select whether to set the offset for DT or for PPM or if you directly want
to set the wave length shift (
∆
WL Offset).
When proceeding moisture measurement in liquids you cannot set the DT
Offset.
● Under “New Setpoint“ enter the externally measured reference value for
the displayed value resp. the direct wave length shift.
The display indicates the current measured values, the date of the last
offset correction and the conditions prevailing at that time (TT and SP) as
well as the wave length shift (
Δ
WL).
Cancel offset
If you want to cancel an offset setting, touch the
Set ΔWL to zero
field. The
wave length shift is reset to zero, consequently no offset is set any more.
Содержание HYGROPHIL F 5673
Страница 2: ......
Страница 15: ...System description 1 9 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 28: ...System description 1 22 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 32: ...Safety precautions 2 4 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 43: ...Installation 3 11 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 110: ...Error handling 6 4 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 122: ...Bus systems 8 10 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 128: ......
Страница 129: ...HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...