Bus systems
8-2
HYGROPHIL® F 5673 Operating manual, Software version 1.8.x, 387896MDHEN(14.06.2013)
The MODBUS slave address has a range from 1 to 247 and could be easily
changed by means of the application software. On the same way the baud
rate (1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200) and the inter-
face’s parity (even, odd, none) can be changed (see section 5.6.7).
The HYGROPHIL
®
F 5673 MODBUS supports function codes „
01
Read
single coils
“ and „
03
Read multiple holding registers
“
Function code
„01“
supplies the master system with the data listed in
Table 1.
Request Host
Function code
1 Byte
0x01
Start Address
2 Bytes
0 to 60
Count of bits
2 Bytes
1 to 61
Response HYGROPHIL
®
F
Function code
1 Byte
0x01
Number of bytes
1 Byte
Number of bytes for the bit information to
be transferred
Bit information
data (bitwise coded) corresponding to the specified
number of bytes
Table 1
01 Read Single Coils
0
CH1 Present?
30
Is Sample CO2 CH1 Valid?
1
CH2 Present?
35
Is Wavelength CH2 Valid?
2
CH3 Present?
36
Is Volume percent CH2 Valid?
3
HCDT HW Present?
37
Is PPM CH2 Valid?
4
Error Status CH1
38
Is Dew Point CH2 Valid?
5
Limit Status CH1
39
Is Frost Point CH2 Valid?
6
Liquid measurement CH1
40
Is Vapour pressure CH2 Valid?
7
Error Status CH2
41
Is Moisture content CH2 Valid?
8
Limit Status CH2
42
Is Sample Temperature CH2 Valid?
9
Liquid measurement CH2
43
Is Sample Pressure CH2 Valid?
10
Error Status CH3
44
Is Sample CO2 CH2 Valid?
11
Limit Status CH3
49
Is Wavelength CH3 Valid?
12
Liquid measurement CH3
50
Is Volume percent CH3 Valid?
13
HCDT Error
51
Is PPM CH3 Valid?
21
Is Wavelength CH1 Valid?
52
Is Dew Point CH3 Valid?
22
Is Volume percent CH1 Valid?
53
Is Frost Point CH3 Valid?
23
Is PPM CH1 Valid?
54
Is Vapour pressure CH3 Valid?
24
Is Dew Point CH1 Valid?
55
Is Moisture content CH3 Valid?
25
Is Frost Point CH1 Valid?
56
Is Sample Temperature CH3 Valid?
26
Is Vapour pressure CH1 Valid?
57
Is Sample Pressure CH3 Valid?
27
Is Moisture content CH1 Valid?
58
Is Sample CO2 CH3 Valid?
28
Is Sample Temperature CH1 Valid?
61
Is Hydro Carbon Dew Point Valid?
29
Is Sample Pressure CH1 Valid?
Содержание HYGROPHIL F 5673
Страница 2: ......
Страница 15: ...System description 1 9 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 28: ...System description 1 22 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 32: ...Safety precautions 2 4 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 43: ...Installation 3 11 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 110: ...Error handling 6 4 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 122: ...Bus systems 8 10 HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...
Страница 128: ......
Страница 129: ...HYGROPHIL F 5673 Operating manual Software version 1 8 x 387896MDHEN 14 06 2013 ...