If no parameter is changed, the internal date / time on the pressure gauge is not modified.
During the date / time update, the seconds are automatically set to zero.
23. OPERATION of the DATA LOGGER
The
DataLogger
allows the storage of up to
130,000
(
65,000
if the temperature is also stored) measurement
points in steps ranging from 1s to 10 hours according to the parameter
t1
defined in the
Parameter Menu
. The
duration of the cycle is established by the parameter
t2
always defined in the
Parameter menu
.
Each reading cycle can be stopped before the set time and restarted with the same settings up to four times,
or it is possible to repeat the last performed cycle once again up to four times. Once the limits is reached, it is
no longer possible to continue and a new cycle must be started with consequent loss of the recorded data. If,
during the repetition of the cycle, the maximum recording capacity is reached, the cycle stops.
WARNING
! The initialization of a new cycle causes the
loss of previously stored data.
For very long Datalog cycles it is possible to save on battery consumption by using the stand-by function of the
pressure gauge.
This feature is activated automatically when the acquisition time between 2 storage points is greater than the
Auto Power Off time.
Example:
Auto Power Off Time: 1 minute
Acquisition time (t1): 5 minutes
In these conditions, in the interval between the two acquisitions, the pressure gauge switches off the display
and deactivates the other functions. It is reactivated 30 seconds seconds before making a measurement and
for the next 5 seconds.
In any case, during a datalog cycle, the auto switch off function of the pressure gauge is not active.
If an excessively low battery level is detected, the datalog cycle will automatically stop. All data stored
up to that point are saved in the internal memory.
For particularly long datalog cycles, the internal battery on the pressure gauge is not sufficient. The
pressure gauge must be powered externally through the USB port.
Note:
The functionality of the DataLogger is fully manageable by PC using the
Quick Analyzer
software through which
it is possible to:
Содержание LABDMM2
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 18: ......
Страница 27: ...LABDMM2 Manuale Operativo MO LABDMM2 560 R0 Pag 26 33 Dimensioni mm CONTENITORE DA INCASSO...
Страница 28: ...MO LABDMM2 560 ENG R1 PROFESSIONAL DIGITAL MANOMETER for PRESSURE and TEMPERATURE measurement...
Страница 57: ...LABDMM2 USER GUIDE MO LABDMM2 560 ENG R1 Pag 29 34 Dimensions mm BUILT IN VERSION...