LABDMM
2
USER GUIDE
MO.LABDMM2.560.ENG.R1
Pag. 21
29. TEMPERATURE CALIBRATION
The pressure gauge has an internal temperature sensor with an accuracy of 1 ° C and a resolution of 0.1 ° C.
If the temperature reading does not fall within the specifications or if you want to improve this performance,
it is possible to calibrate the sensor.
Calibration of the temperature sensor is performed at two different temperature points.
The first point at room temperature (
t1
) the second at a temperature of about 40 or 50 degrees ° C (
t2
).
+
Press the keys simultaneously for a few seconds
SET
and
PEAK.
P0000
Set the password
3126
(
) and confirm with
SET.
t1
The message
t1
will be displayed to indicate that the room temperature can be
calibrated.
Confirm with
SET
.
25.3°C
The display shows the temperature read by the pressure gauge.
With a reference thermometer, measure the ambient temperature.
Correct the value displayed with the
keys until it is the same as the reference
measurement.
Confirm with
SET
.
t2
The message
t2
will be displayed to indicate that the second temperature point
can be calibrated.
Bring the pressure gauge to the desired temperature and wait until the system is
stabilized.
Confirm with
SET
.
51.3 °C
The display shows the temperature read by the pressure gauge.
With a reference thermometer measure the temperature of the chamber.
Correct the value displayed with the
keys until it is the same as the reference
measurement.
Confirm with
SET
.
End
The procedure is complete.
In the event that problems should arise in the calibration procedure, it is possible to return to the factory
calibration using the password
3125
.
Содержание LABDMM2
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 18: ......
Страница 27: ...LABDMM2 Manuale Operativo MO LABDMM2 560 R0 Pag 26 33 Dimensioni mm CONTENITORE DA INCASSO...
Страница 28: ...MO LABDMM2 560 ENG R1 PROFESSIONAL DIGITAL MANOMETER for PRESSURE and TEMPERATURE measurement...
Страница 57: ...LABDMM2 USER GUIDE MO LABDMM2 560 ENG R1 Pag 29 34 Dimensions mm BUILT IN VERSION...