Operation
4
Ope ration
4.1
Putting into service
If you are putting a new or repaired device into service, you should first
select a suitable location for it
connect the necessary cords
and fill the device with heat transfer fluid.
CAUTION
You must disinfect the heating cushions and heating chamber before using the de-
vice. This procedure is described in the Chapter Cleaning of these instructions.
4.1.1
Siting the device
The Barkey plasmatherm is designed for use as a fixed installation inside buildings. It is
not intended for mobile use.
The Barkey plasmatherm must be set up on a stable, hard and level surface. Unobstructed
access to the device from above and in front must be guaranteed.
CAUTION
The device must be lifted by a minimum of 2 people. When carrying, grip the de-
vice by the bottom edge only. Recesses are provided in the base for safe carrying.
CAUTION
The device is ventilated from below. It should therefore
not
be placed on a soft sur-
face into which its feet could sink. There must be a minimum distance of 50 mm on
3 sides of the Barkey plasmatherm between it and walls, cabinets or other devices.
CAUTION
When selecting the location for the device, it is imperative that you maintain the
distances stated in the Chapter
of this
instruction manual from other devices with electromagnetic emission.