1
InstructionsForUse
9TechnicalSpecifications
9.1
OperatingandStorageConditions
Operating Temperature:
59°F to 100°F (15°C to 38°C)
Storage Temperature:
50°F to 95°F (10°C to 35°C)
Operating Humidity:
5% to 85% Relative Humidity (non-condensing)
Storage Humidity (packaged):
5% to 95% Relative Humidity (non-condensing)
Storage Humidity (unpackaged):
5% to 85% Relative Humidity (non-condensing)
9.
ScannerSpecifications
Dimensions:
12" W x 12" H x 5" D
Weight:
10 lbs.
Power Sources:
AC adapter, Internal and External DC Battery Pack
Power Consumption:
82 Watts Maximum
Monitor Size:
12.1” diagonal
IEC 60601- 1:
Class II, Type BF Applied Part, Continuous Operation, Internally Powered
Equipment, Not Category AP or APG Equipment, Not protected against
ingress of water.
9.
ProbeAcousticOutputSpecifications
Description
of Probe
Operating
Mode
Ispta.X
(X denotes statisti-
cally determined
maximum)
FDA I-spta.3
Published
Values
MI X
(X denotes
statistically
determined
maximum)
FDA MI
Published Values
L-VA
B
78.2mW/cm2
Peripheral Vessel <
720 m W/cm2
Cardiac <
430 mW/cm2
Fetal Imaging & Other** <
94 mW/cm2
1.1
Peripheral Vessel <
1.9
Cardiac <
1.9
Fetal Imaging & Other** <
1.9