0
Site~Rite*5
Ultrasound System
.6
ConnectingandDisconnectingSite~Rite
*
5Probes
.7
PoweringontheSite~Rite
*
5UltrasoundSystem
To power on the Site~Rite
*
5 Ultrasound System
1. Verify that the probe is connected to the Site~Rite 5 scanner.
2. Depress and release the power button located on the front of the Site-Rite 5 scanner.
3. Wait approximately 20 seconds for the display screen to illuminate.
4. To power off the Site~Rite 5 System, depress and release the power button again. The
system is powered off when the display screen darkens.
ProbeConnected
USBConnected
Device
Warning:
Use only Bard Access Systems
probes with this system. Use of unapproved
probes may result in patient injury or equip-
ment damage.
Caution:
Do not pull the cable to discon-
nect the probe connector from the scanner.
Pulling the cable may damage the cable,
cable connection or scanner.
Caution:
Do not twist or bend the probe
cable in excess of that required during
normal use of the probe. Excessive twisting
or bending of the cable may cause failure,
intermittent or unpredictable operation.