Sherlock 3CG™ Tip Confirmation System
Page 30
Magnetic navigation Stylet icon has erratic
behavior
Cause
: Stylet at edge of sensor’s depth range (3
to 11 cm).
Solution
:
1.
Pull catheter back until stylet icon is
stable.
2.
Verify catheter is moving in the correct
direction and continue placement.
Cause
: Magnetic interference.
Solution
: Pull catheter tip back at least 12 inches
(30 cm) from the sensor, recalibrate, and re-
advance catheter.
Cause
: Too much time has passed since
calibration. Magnetic fields have changed.
Solution
: Pull catheter tip back at least 12 inches
(30 cm) from the sensor, recalibrate and re-
advance catheter.
Magnetic navigation Stylet icon disappears,
"sweeping magnifying glass" displayed.
Cause
: Stylet is out of sensor’s depth range (3 to
11 cm).
Solution
:
1.
Pull catheter back until stylet icon is
stable.
2.
Verify catheter is moving in the correct
direction and continue placement using
ECG guidance and catheter
measurement.
Cause
: Magnetic interference.
Solution
: Pull catheter tip back at least 12 inches
(30 cm) from the sensor, recalibrate, and re-
advance catheter.
Cause
: Too much time has passed since
calibration. Magnetic fields have changed.
Solution
: Pull catheter tip back at least 12 inches
(30 cm) from the sensor, recalibrate and re-
advance catheter.
If problems continue, call the technical support hotline at 1(800) 443-3385.