Sherlock 3CG™ Tip Confirmation System
Page 18
SHERLOCK 3CG™ TCS CATHETER CARE GUIDANCE
Step 1: Prepare the device.
Connect the sensor to the Site~Rite Vision™ II Ultrasound System via USB cord.
Verify that the external ECG flatline signal is scrolling.
Verify the battery charge is sufficient for the procedure.
Enter Patient Information as needed.
Step 2: Position the patient and perform ultrasound pre-scan.
Refer to the Bard Access Systems’ Site~Rite Vision™ II Ultrasound System Instructions for
Use.
Step 3: Measure the catheter length.
Refer to Bard Access Systems’ Catheter Instructions for Use.
Step 4: Prepare the sensor.
Slide the Fin Assembly onto the sensor until fully seated.
Place sensor in the sensor holder with the fin and ECG electrodes remaining outside the
holder and tighten the cinch ring.
1
2
3
4