59
R5913964 /00
MDSC-8255
Immunity test
IEC 60601-1-2 4
th
edition
(2014)
Test levels
Compliance level
Electromagnetic
environment
–
guidance
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V/m (150 kHz to 80
MHz)
9 to 28 V/m in
communication service
channels up to 6 GHz
3 V/m (150 kHz to 80
MHz)
9 to 28 V/m in
communication service
channels up to 6 GHz
Portable and mobile RF
communications
equipment should be used
no closer to any part of the
MDSC-8255, including
cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended separation
distance
d = 1.2
√
P
d = 1.2
√
P
80 MHz to 800
MHz
d = 2.3
√
P
800 MHz to 2.5
Ghz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the recommended
separation distance in
meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
survey,
2
should be less
than the compliance level
in each frequency range.
3
Interference may occur in
the vicinity of equipment
marked with symbol:
At 80 MHz and 800 MHz, the higher frequency range applies.
2.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the MDSC-8255 is used exceeds the applicable RF compliance level above, the MDSC-8255
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the MDSC-8255.
3.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Содержание MDSC-8255 12G
Страница 6: ...R5913964 00 MDSC 8255 6 ...
Страница 7: ...7 R5913964 00 MDSC 8255 Welcome 1 ...
Страница 10: ...R5913964 00 MDSC 8255 10 Welcome ...
Страница 11: ...11 R5913964 00 MDSC 8255 Parts controls and connectors 2 ...
Страница 18: ...R5913964 00 MDSC 8255 18 Parts controls and connectors ...
Страница 19: ...19 R5913964 00 MDSC 8255 Installation 3 ...
Страница 26: ...R5913964 00 MDSC 8255 26 Installation ...
Страница 27: ...27 R5913964 00 MDSC 8255 Daily operation 4 ...
Страница 32: ...R5913964 00 MDSC 8255 32 Daily operation ...
Страница 49: ...49 R5913964 00 MDSC 8255 Important information 6 ...
Страница 70: ...R5913964 00 MDSC 8255 70 Important information ...
Страница 71: ......