Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The MDSC-8255 is intended for use in the electromagnetic environment (IEC 60601-1-2 4
th
edition) specified
below. The customer or the user of the MDSC-8255 should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 4
th
edition
(2014)
Test levels
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8kV contact
± 15kV air
± 8kV contact
± 15kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 2kV for power supply
lines
± 1kV for input/ output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
< 5% U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user of
the MDSC-8255 requires
continued operation
during power mains
interruptions, it is
recommended that the
MDSC-8255 be powered
from an uninterruptible
power supply or a battery.
1.
is the a.c. mains voltage prior to application of the test level.
Содержание MDSC-8255 12G
Страница 6: ...R5913964 00 MDSC 8255 6 ...
Страница 7: ...7 R5913964 00 MDSC 8255 Welcome 1 ...
Страница 10: ...R5913964 00 MDSC 8255 10 Welcome ...
Страница 11: ...11 R5913964 00 MDSC 8255 Parts controls and connectors 2 ...
Страница 18: ...R5913964 00 MDSC 8255 18 Parts controls and connectors ...
Страница 19: ...19 R5913964 00 MDSC 8255 Installation 3 ...
Страница 26: ...R5913964 00 MDSC 8255 26 Installation ...
Страница 27: ...27 R5913964 00 MDSC 8255 Daily operation 4 ...
Страница 32: ...R5913964 00 MDSC 8255 32 Daily operation ...
Страница 49: ...49 R5913964 00 MDSC 8255 Important information 6 ...
Страница 70: ...R5913964 00 MDSC 8255 70 Important information ...
Страница 71: ......