8. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood,
concrete or ceramic tile.
If
fl
oors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 2kV for power supply
lines
± 1kV for input/ output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
< 5% U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
< 5% U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
Mains power quality
should by that of a
typical commercial or
hospital environment. If
the user of the Coronis
Fusion 4MP / 6MP
requires continued
operation during power
mains interruptions, it is
recommended that the
Coronis Fusion 4MP /
6MP be powered from
an uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic
fi
eld
IEC 61000-4-8
3 A/m
Not applicable
2
Power frequency
magnetic
fi
elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Coronis
Fusion 4MP / 6MP,
including cables, than
the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d = 1.2
√
P
d = 1.2
√
P
80 MHz to 800
MHz
1. is the a.c. mains voltage prior to application of the test level.
2. Coronis Fusion 4MP / 6MP doesn’t contain susceptible components to magnetic fields
K5902056 CORONIS FUSION 4MP / 6MP 31/10/2014
45
Содержание MDCC-4230
Страница 1: ...Coronis Fusion 4MP 6MP User Guide MDCC 4230 MDCC 6330 K5902056 00 31 10 2014 ...
Страница 6: ...1 Welcome 4 K5902056 CORONIS FUSION 4MP 6MP 31 10 2014 ...
Страница 18: ...3 Display installation 16 K5902056 CORONIS FUSION 4MP 6MP 31 10 2014 ...
Страница 22: ...4 Daily operation Cancel Standby IEC 60417 5009 20 K5902056 CORONIS FUSION 4MP 6MP 31 10 2014 ...
Страница 36: ...5 Advanced operation 34 K5902056 CORONIS FUSION 4MP 6MP 31 10 2014 ...
Страница 38: ...6 Cleaning your display 36 K5902056 CORONIS FUSION 4MP 6MP 31 10 2014 ...