background image

14 

Specifications

 

 

Specifications 

 

Model 

TB200 

Measurement Method 

ISO 7027 nephelometric method (90° ) 

Range 

0

 

~

 

2000 NTU/FNU, 0

 

~

 

500 EBC, 0

 

~

 

9999 ASBC, TSS range deponds on conversion factor 

Resolution 

0.01 (0

 

~

 

100 NTU), 0.1 (100

 

~

 

999 NTU), 1 (999

 

~

 

2000 NTU) 

Accuracy 

± 2% (0

 

~

 

500 NTU), ± 3% (501

 

~

 

2000 NTU) 

Calibration Standards 

0.02, 10.00, 200, 500, 1000, 1500, 2000 NTU 

Light Source 

Infrared-emitting diode 

Detector 

Silicon photodiode 

Stray Light 

< 0.02 NTU 

Sample Vial 

60

 

(H) ×  25

 

(Dia)

 

mm 

Memory 

Stores up to 200 data sets 

Output 

USB communication interface 

Display 

3.5" TFT display 

Operating Temperature 

0

 

~

 

50° C, 32

 

~

 

122° F 

Power Requirements 

DC12V/2A, using AC adapters, 220V/50Hz 

Dimensions 

250

 

(L) ×  177

 

(W) ×  96

 

(H)

 

mm 

Weight 

1.2kg 

 

Optional Accessories 

 

Order Code 

Description 

TB-GV 

Sample vial, glass body, 60

 

(H) ×  25

 

(Dia)

 

mm 

USB-BC 

USB cable, 1 meter, USB-B plug 

DCPA-12V 

DC12V power adapter, 1 meter cable, european standard plug 

 
 
 

 
 
 
 
 
 
 
 
 
 
 
 

Содержание TB200

Страница 1: ...TB200 Laboratory Turbidity Meter Instruction Manual Bante Instruments Inc...

Страница 2: ...bration 7 Turbidity measurement 9 Low turbidity samples 200 NTU 9 High turbidity samples 2000 NTU 9 Total suspended solids measurement 10 Calculating the TSS factor 10 Setting the TSS factor 10 TSS me...

Страница 3: ...ent work environment meets the following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C 32 F and less than 50 C 122 F No potential magnetic fields and ambient lig...

Страница 4: ...oll down through the menu item Confirms the calibration settings or displayed options Starts measurement Locks the measured value Switching the Meter On and Off Insert the connector of power adapter i...

Страница 5: ...0 1 0 01 TSS Factor Set the conversion factor for measurement of the total suspended solids 0 13 Auto Power Off 2 hours When the option is enabled the meter will automatically switch off if no key is...

Страница 6: ...libration or settings 3 1 Press and hold the key for 3 seconds to enter the setup menu 3 2 Press the key to select the Password press the Enter key to confirm 3 3 Press the key to select the Enable Pr...

Страница 7: ...brated 4 1 Press and hold the key for 3 seconds to enter the setup menu 4 2 Press the key to select the Reset press the Enter key to confirm 4 3 Press the key to select Restore factory settings Press...

Страница 8: ...ause inaccurate readings Turbidity Calibration The TB200 turbidity meter is calibrated with Formazin Standards at the factory and does not require user calibration prior to use Do not shake or agitate...

Страница 9: ...th the arrow on the meter Close the light shield lid 6 Press the Enter key the meter begins the calibration The Calibrating icon shows on the upper right corner of the screen 7 Wait for the reading to...

Страница 10: ...ops and fingerprints Ensure that the outside of vial is dry and clean 1 4 Place the vial in the measurement chamber Align the mark on the vial with the arrow on the meter 1 5 Close the light shield li...

Страница 11: ...for them decrease to room temperature 1 8 Weigh the filter disk and watch glass and record the mg value as A 1 9 Calculate the TSS value using the formula below TSS mg L 1 10 Fill the vial with the s...

Страница 12: ...y to switch pages 3 Press the ESC key the meter returns to the measurement mode Deleting the data log If the memory is full the meter will automatically show a reminder screen when the MI key is press...

Страница 13: ...uter 2 Click the OK button the application starts If your computer can not find a communication port please click the file CP210xVCPInstaller_x64 exe or CP210xVCPInstaller_x86 exe to update the drive...

Страница 14: ...urbidity free water The turbidity free water is used for preparation of turbidity standards and is prepared by filtering deionized water through a 0 45 m or smaller pore sized membrane Preparation of...

Страница 15: ...ight Source Infrared emitting diode Detector Silicon photodiode Stray Light 0 02 NTU Sample Vial 60 H 25 Dia mm Memory Stores up to 200 data sets Output USB communication interface Display 3 5 TFT dis...

Страница 16: ...r suggestions on how to speed up these efforts Warranty The warranty period for meter is two years from the date of shipment Above warranty does not cover the calibration standards Out of warranty pro...

Отзывы: