Measurement and Calibration

 

 

3. 

Take out the corresponding calibration standard from the accessory box. Wipe the vial with a lint-free cloth to remove fingerprints. 

4. 

Hold the vial cap and gently inverting vial several times. Ensure that the calibration standard is homogeneous and no air bubbles. 

5. 

Insert the calibration standard into the measurement chamber, the mark on the vial cap aligns with the arrow on the meter. Close the light   

 

shield lid. 

 

   

   

 

 
6. 

Press the 

Enter

 key, the meter begins the calibration. The 

Calibrating

” 

icon shows on the upper right corner of the screen. 

7. 

Wait for the reading to stabilize, the meter automatically shows the next calibration point. 

 

   

 

 
8. 

If necessary, press the 

 key to select an applicable calibration point. 

9. 

Repeat steps 3 through 6 until the screen shows 

“Calibration is Completed”

. The meter returns to the measurement mode automatically. 

 

Exit the calibration 

To exit the calibration mode without saving the calibration, press the 

ESC

 key. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Clean 

Mix 

Align arrows 

Содержание TB200

Страница 1: ...TB200 Laboratory Turbidity Meter Instruction Manual Bante Instruments Inc...

Страница 2: ...bration 7 Turbidity measurement 9 Low turbidity samples 200 NTU 9 High turbidity samples 2000 NTU 9 Total suspended solids measurement 10 Calculating the TSS factor 10 Setting the TSS factor 10 TSS me...

Страница 3: ...ent work environment meets the following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C 32 F and less than 50 C 122 F No potential magnetic fields and ambient lig...

Страница 4: ...oll down through the menu item Confirms the calibration settings or displayed options Starts measurement Locks the measured value Switching the Meter On and Off Insert the connector of power adapter i...

Страница 5: ...0 1 0 01 TSS Factor Set the conversion factor for measurement of the total suspended solids 0 13 Auto Power Off 2 hours When the option is enabled the meter will automatically switch off if no key is...

Страница 6: ...libration or settings 3 1 Press and hold the key for 3 seconds to enter the setup menu 3 2 Press the key to select the Password press the Enter key to confirm 3 3 Press the key to select the Enable Pr...

Страница 7: ...brated 4 1 Press and hold the key for 3 seconds to enter the setup menu 4 2 Press the key to select the Reset press the Enter key to confirm 4 3 Press the key to select Restore factory settings Press...

Страница 8: ...ause inaccurate readings Turbidity Calibration The TB200 turbidity meter is calibrated with Formazin Standards at the factory and does not require user calibration prior to use Do not shake or agitate...

Страница 9: ...th the arrow on the meter Close the light shield lid 6 Press the Enter key the meter begins the calibration The Calibrating icon shows on the upper right corner of the screen 7 Wait for the reading to...

Страница 10: ...ops and fingerprints Ensure that the outside of vial is dry and clean 1 4 Place the vial in the measurement chamber Align the mark on the vial with the arrow on the meter 1 5 Close the light shield li...

Страница 11: ...for them decrease to room temperature 1 8 Weigh the filter disk and watch glass and record the mg value as A 1 9 Calculate the TSS value using the formula below TSS mg L 1 10 Fill the vial with the s...

Страница 12: ...y to switch pages 3 Press the ESC key the meter returns to the measurement mode Deleting the data log If the memory is full the meter will automatically show a reminder screen when the MI key is press...

Страница 13: ...uter 2 Click the OK button the application starts If your computer can not find a communication port please click the file CP210xVCPInstaller_x64 exe or CP210xVCPInstaller_x86 exe to update the drive...

Страница 14: ...urbidity free water The turbidity free water is used for preparation of turbidity standards and is prepared by filtering deionized water through a 0 45 m or smaller pore sized membrane Preparation of...

Страница 15: ...ight Source Infrared emitting diode Detector Silicon photodiode Stray Light 0 02 NTU Sample Vial 60 H 25 Dia mm Memory Stores up to 200 data sets Output USB communication interface Display 3 5 TFT dis...

Страница 16: ...r suggestions on how to speed up these efforts Warranty The warranty period for meter is two years from the date of shipment Above warranty does not cover the calibration standards Out of warranty pro...

Отзывы: