NOTE: When running a Match inspection, the sensor will look for any possible patterns to match
anywhere within the Field of View.
d. Click the Teach icon to teach the sensor this good reference part. With Annotations Enabled, the screen will highlight in
green the pattern found.
2. Set inspection parameters.
• Adjust the Percent Match parameter. The Percent Match setting adjusts for how closely the inspected part or label matches
the reference part or label. The Percent Match scale is from 0 to 100 where 0 is the most tolerant and 100 is the least tolerant.
Move the slider to the left or to the right.
NOTE: When running a Match inspection with annotations enabled, the sensor will highlight in green
any pattern matches that meet or exceed the value specified for Percent Match. Patterns that are
below the specified value for Percent Match (down to approximately 20%), or out of the Rotation
Range (see below), will be colored yellow
1. Using a "bad" part, click the Manual Trigger icon in the lower-right of the screen to capture an image. For this example,
one of the stored images is missing the letter "N," yet the sensor initially sees this as a "good" label.
2. Go to Main Menu > Inspection , and click the Percent Match option.
iVu Series Image Sensor
Rev. E
www.bannerengineering.com - tel: 763-544-3164
51