66 / 86
6416h GB/2021-01
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
Checking the rinsing and/or moving function
The most complex cleaning program (“Robotic”) is to be used for checking both
functions.
Turn on the ultrasonic bath.
Fill the ultrasonic oscillating tank with liquid.
Select the “Robotic” program and connect the TRISON Twist to the control unit.
Continue pressing the “Next” key until the channel selector appears (specify 8 mm
instrument diameters). Select all channels and start. Skip the following soaking
phase by pressing the “Start” key. Allow the rinsing program to run for approx. 5
min. During this time, check the following functions:
Rinsing function:
On hoses A1, B1, etc., water should run out visibly during the return cycle. A
fl
ow-
through of approx. 350 ml/min should be displayed.
Moving function:
Movement must be visible on the TRISON Twist moving device at the slots. If needed,
connect a test instrument to make the movement recognisable.
6.3
Malfunction/fault analysis
TRISON ultrasonic baths are designed for high reliability. A possible malfunction due to
a defective component cannot be fully ruled out, however. Mechanical defects, e.g. on
the adapters, hoses, etc., are also possible due to wear or improper use.
The following overview of possible sources of error should serve as an aid for error
detection and troubleshooting.
Error
Possible cause(s)
Measure(s)
Instruments do not
move uniformly
The movement has
been programmed to be
intentionally chaotic.
The actuating force is limited,
thus sluggish instruments can
only be moved to a limited
extent.
None
The TRISON Base
cannot be turned on
(display remains dark)
TRISON Base connected to
the mains?
Check the mains connection, see the
installation instructions.
Mains switch correctly
activated?
Check that the mains switch is lit up.
Fuses defective?
If needed, Replace fuses, see
Section 6.4.3.
No displays or
incomprehensible
displays
TRISON Base defective?
Can only be recti
fi
ed by the manufacturer.
Содержание Trison 4000
Страница 80: ...B Maintenance lists Maintenance list daily Replace filter cartridge on the TRISON Base Date Signature ...
Страница 82: ...C Screen displays Sequence without interruptions ...
Страница 83: ......
Страница 84: ......