6416h GB/2021-01
65 / 86
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
6.2
Function checks
Checking the touchscreen
Turn on the ultrasonic bath – the touchscreen must light up.
The touchscreen should respond to the pushing of buttons and the relevant screens
must appear.
Checking operation of the ultrasonic bath
The output can be checked using a standard wattmeter. It is to be inserted between the
mains plug, the mains supply switch, and the power outlet.
Turn on the ultrasonic bath.
Fill the ultrasonic oscillating tank with liquid.
Only the ultrasound needs to be turned on for this check. Select
“Standard” and start.
Compare the value displayed with the corresponding value in the
technical data (Section 1.8) (tolerances ± 10 %).
Check the ultrasound e
ff
ect
A foil test is recommended for veri
fi
cation.
A suitable foil test frame is included in the scope of delivery. Household aluminium foil
is used to conduct the test. Next, a comparison is made with previously-generated foils.
As a departure from the detailed information in the annex (which is intended for
simple standard ultrasonic baths), please note the following special instructions for the
TRISON:
After turning on the device, the test can be conducted by going to
Settings/Service/Foil
test
.
Important:
After the foil test, the ultrasonic oscillating tank must be thoroughly rinsed so that no
aluminium residue remains on the inside of the TRISON Base.
Note:
A measurement procedure is described in DIN SPEC 40170:2013-11 (measurement
and evaluation of the cavitation noise).